User`s manual
WaveGen – built-in function generator (specications are typical)
Waveforms Sine, square, pulse, triangle, ramp, noise, DC
Sine – Frequency range: 0.1 Hz to 20 MHz
– Amplitude atness: ±0.5 dB (relative to 1 kHz)
– Harmonic distortion: –40 dBc
– Spurious (non harmonics): –40 dBc
– Total harmonic distortion: 1%
– SNR (50 ohm load, 500 MHz BW) : 40 dB (Vpp ≥ 0.1 V); 30 dB (Vpp <0.1 V)
Square wave/pulse – Frequency range: 0.1 Hz to 10 MHz
– Duty cycle: 20 to 80%
– Duty cycle resolution: Larger of 1% or 10 ns
– Pulse width: 20 ns minimum
– Pulse width resolution: 10 ns or 5 digits, whichever is larger
– Rise/fall time: 18 ns (10 to 90%)
– Overshoot: < 2%
– Asymmetry (at 50% DC): ±1% ± 5 ns
– Jitter (TIE RMS): 500 ps
Ramp/triangle wave – Frequency range: 0.1 Hz to 100 kHz
– Linearity: 1%
– Variable symmetry: 0 to 100%
– Symmetry resolution: 1%
Noise Bandwidth: 20 MHz typical
Frequency – Sine wave and ramp accuracy:
– 130 ppm (frequency < 10 kHz)
– 50 ppm (frequency > 10 kHz)
– Square wave and pulse accuracy:
– [50+frequency/200] ppm (frequency < 25 kHz)
– 50 ppm (frequency ≥ 25 kHz)
– Resolution: 0.1 Hz or 4 digits, whichever is larger
Amplitude – Range:
– 20 mVpp to 5 Vpp into Hi-Z
– 10 mVpp to 2.5 Vpp into 50 ohms
– Resolution: 100 µV or 3 digits, whichever is larger
– Accuracy: 2% (frequency = 1 kHz)
DC offset – Range:
– ±2.5 V into Hi-Z
– ±1.25 V into 50 ohms
– Resolution: 100 μV or 3 digits, whichever is larger
– Accuracy: ±1.5% of offset setting ±1.5% of amplitude ±1 mV
Trigger output Trigger output available on Trig out BNC
Performance characteristics (continued)
18 | Keysight | InniiVision 2000 X-Series Oscilloscopes - Data Sheet