User`s manual

Waveform math
Operators Add, subtract, multiply, FFT
FFT Windows: Hanning, at top, rectangular; Blackman-Harris - up to 64 kpts resolution
Sources Math functions available between any two channels
Display characteristics
Display 8.5-inch WVGA
Resolution 800 (H) x 480 (V) pixel format (screen area)
Interpolation Sin(x)/x interpolation (using FIR lter; used when there is less than one sample per column of the display)
Persistence Off, innite, variable persistence (100ms-60s)
Intensity gradation 64 intensity levels
Modes Normal
XY - XY mode changes the display from voltage versus time scale to a volts versus volts scale
Roll – Displays the waveform moving across the screen from right to left much like a strip chart recorder
MSO (digital channels)
Upgradable from DSO Yes
MSO channels 8 channels (D0 to D7)
Maximum sample rate 1 GSa/s
Maximum record length 500 kpts per channel (digital channels only)
125 kpts per channel (analog and digital channels)
Threshold selections TTL (+1.4 V), CMOS (+2.5 V), ECL (-1.3 V), User-denable (±8.0 V in 10 mV stops)
Threshold accuracy* ± (100 mV + 3% of threshold settings)
Maximum input dynamic
range
±10 V about threshold
Minimum voltage swing 500 mVpp
Input impedance 100 kΩ ± 2% at probe tip, ~8 pF
Minimum detectable pulse
width
5 ns
Channel-to-channel skew 2 ns (typical), 3 ns (maximum)
Performance characteristics (continued)
* Denotes warranted specications, all others are typical.
Specications are valid after a 30-minute warm-up period and from ±10 °C rmware calibration temperature.
** 1 mV/div and 2 mV/div is a magnication of 4 mV/div setting. For vertical accuracy calculations, use full scale of 32 mV for
2 mV/div sensitivity setting.
Cursors
Types Amplitude, time , frequency (FFT), manual, tracking, binary, HEX
Measurements ΔT, 1/ΔT, ΔV/X, 1/ΔX, ΔY, Phase and Ratio
Cursors** Single cursor accuracy:
±[DC vertical gain accuracy + DC vertical offset accuracy + 0.25% full scale]
Dual cursor accuracy:
±[DC vertical gain accuracy + 0.5% full scale]*
Automatic waveforms measurements
Voltage Snapshot all, maximum, minimum, peak-to-peak, top, base, amplitude, overshoot, preshoot, average- N
cycles, average-full screen, DC RMS- N cycles, DC RMS- full screen, AC RMS- N cycles
Time
Period, frequency, rise time, fall time, + width, – width, duty cycle, delay AB (rising edge),
delay AB (falling edge), phase AB (rising edge,) and phase AB (falling edge)
17 | Keysight | InniiVision 2000 X-Series Oscilloscopes - Data Sheet