User`s guide

2 General Power Meter Functions
48 N1911A/1912A P-Series Power Meters User’s Guide
Enter (or edit) the frequency and offset pairs as follows:
1 Press to add a new frequency value (or press to edit).
Use the numeric keypad to enter the required value in the Frequency
pop- up window. Complete the entry by pressing the , keys.
2 Enter the new offset value (or press to edit). Use the numeric
keypad to enter the required value in the Offset pop- up window.
Complete the entry by pressing the key.
3 Continue adding/editing values until you have entered all the data you
require.
4 When you have finished editing the table press to save the table.
Insert
Change
GHz
MHz
Change
%
Done
NOTE
If you measure a signal with a frequency outside the frequency range defined in the
frequency dependent offset table, the power meter uses the highest or lowest frequency
point in the frequency dependent offset table to calculate the offset