User`s guide

Introduction 1
N1911A/1912A P-Series Power Meters User’s Guide 25
Table 1 - 1 Typ e o f Lan Status message.
Figure 1-12 Example of Lan Status message.
Type of message Description
Lan: No Fault
Status: Initialized
A valid IP address is successfully obtained using selected
LAN configuration and the network state is initialized.
Lan: No Fault
Status: Running
A valid IP address is successfully obtained using selected
LAN configuration, while network is running.
Lan: Fault
Status: Initialization failed
IP conflict occurs, or
IP address failed to be obtained using selected LAN
configuration, or
No LAN configuration is selected.
Lan: Fault
Status: Disconnected
LAN cable is unplugged.
Lan: –
Status: Restarting
Restart the network and try to obtain an IP address using
selected LAN configuration.
Lan: DHCP Not Available
Status: Running
Unable to obtain IP address from DHCP server (if user
select DHCP configuration)
IP address obtained from Auto-IP or manual
configuration.
NOTE
For more details on remote interface configuration, refer to P-Series Power Meters
Installation Guide.