User`s guide
Using E9300 E-Series Power Sensors 5
N1911A/1912A P-Series Power Meters User’s Guide 153
Power Meter Configuration
The P- Series power meters automatically recognize an E- Series E9300
power sensor when it is connected. The sensor calibration data is
automatically read by the power meter. The power meter also configures
the auto- averaging settings shown in Figure 5- 113 to suit the power
sensor characteristics.
Figure 5-113E9300 E-Series auto-averaging settings
-4 dBm
-10 dBm
Sensor Power
Minimum
Sensor Dynamic Range
10 dBm
2 dBm
1234
1114
Resolution Setting
Number of Averages
11416
11
8
32
1416128
11664128
Maximum
-40 dBm
-50 dBm
-20 dBm
-30 dBm
1114
11216
121664
4 16 128 256
32 64 256 256
Upper Range
Lower Range
Sensor Power
6 dBm
0 dBm
20 dBm
12 dBm
-30 dBm
-40 dBm
-10 dBm
-20 dBm
26 dBm
20 dBm
40 dBm
32 dBm
-10 dBm
-20 dBm
10 dBm
0 dBm
E9300/1B
E9300/1H
E9300/1/4A
NOTE
These values are valid only for the power meter channel connected with E-Series E9300
power sensor. You can also configure the settings manually – refer to “Achieving Stable
Results with TDMA Signals” on page 160 if required.