User`s guide

General Power Meter Functions 2
N1911A/1912A P-Series Power Meters User’s Guide 83
Example of SCPI commands set
Refer to the following SCPI commands set for normal mode and average
mode in frequency sweep measurement.
Normal mode
This feature is applicable for E9320 and N1920 Series power sensor when
connected. Below are examples of SCPI commands when used in normal
mode. Send the SCPI commands below to meter using remote interface.
SENS:DET:FUNC NORM
TRIG:SOUR EXT
SENS:FREQ:START 50 MHz
SENS:FREQ:STOP 4 GHz
SENS:FREQ:STEP 2048
Optional measurement configurations:
TRIG:DEL 1ms
TRIG:HOLD 10us
OUTP:TRIG ON
SENS:SWE1:OFFS:TIME 100us
SENS:SWE1:TIME 100us
INIT:CONT ON
*OPC
*ESR?
FETC?
NOTE
Refer to the P-Series Power Meters Programming Guide for the commands usage details.