Technical data

Theory of Operation 4
N1911A/1912A P-Series Power Meters Service Guide 43
Main Board Assembly
Purpose
Provides the average measurement path(s)
Provides the peak measurement path(s) to the DAP(s)
Provides external trigger input/ output and recorder output(s)
Provides the driver and the LVDS serialiser for the LCD display
Provides signal routing between the PPMC, DAP(s), and front panel
Inputs
Power supplies [from PSU]
Sensed power level(s) [from sensor flexi(s)]
Trigger input [from external equipment]
Control, address, and data lines [from PPMC]
Outputs
Processed average path measurement [to PPMC]
Unprocessed peak path measurement samples [to DAP(s)]
Trigger output & recorder output(s) [to external equipment]
LVDS LCD display contorl lines [to front panel]
Control and data lines [to PPMC]