Technical data
3 Adjustments
38 N1911A/1912A P-Series Power Meters Service Guide
Introduction
Attempts to correct the power reference level if the performance test has
failed. Power reference level is controlled by the coarse and fine settings of
a digital potentiometer. Adjustment of the coarse and fine settings can only
be carried out via remote commands. Adjustment can be carried out with-
out having to remove the outer covers from the DUT.