Linearity (Average Path)
Description
Linearity over the full input voltage range of the measurement path is
warranted to provide assurance of instrument accuracy. This test measures
linearity by using a calibration DAC and a calibration ADC (built into the
DUT) to stimulate and compare performance of the average path against
the measurement ADC, returning the worst case percentage error.
Equipment
• No test equipment required
Test Method
1 Execute the internal linearity measurement procedure for channel A.
2 Read back the result of the measurement from the DUT.
3 If the DUT model number is N1912A, then repeat this procedure for
channel B.
4 The test will take a few minutes to complete.
5 The measurement result should be less than 0.5 and greater than –0.5.
The optimum measurement result for this test is 0.
NOTE
• This test can be performed manually via the commands:
SERV:BIST:CW[1|2]:LIN
SERV:BIST:CW[1|2]:LIN:PERR?
(Refer to the programming guide for further details on the use of these commands)
• No adjustment is available for this test if it fails (see Chapter 5, “Troubleshooting
Guide”).