Technical data

Performance Tests 2
N1911A/1912A P-Series Power Meters Service Guide 29
Test Method
1 Enable the path that routes the time base signal to the trigger output
connector.
2 Using the 53132A, measure the frequency of the signal at the trigger
output connector.
NOTE
This test can be configured manually via the command SERV:BIST:TBAS:STAT
ON, which enables the 10 MHz feed to the trigger output connector (refer to the
programming guide for further details on the use of this command).
This test can also be configured manually via the front panel; access the Service menu,
select Self Test, and select Time Base to enable the 10 MHz feed to the trigger output
connector.
No adjustment is available for this test if it fails (see Chapter 5, “Troubleshooting
Guide”).