Technical data
Performance Tests 2
N1911A/1912A P-Series Power Meters Service Guide 23
1 mW Power Reference Level Test
Description
The 1 mW power reference is used for the calibration of 8480 Series, N8480
Series and E- Series power sensors, and is traceable to national standards.
This test uses an 8482A power sensor to transfer the power measured on
an accurately calibrated E4419B or E4417A power meter to the DUT ref-
erence.
Equipment
• Required test equipment:
1 unit of E4419B or E4417A dual channel power meter,
1 unit of 8482A power sensor.
11730A power sensor cable
• Either of these E4419B or E4417A power meters can be used. This
specific power sensor model must be used.
Figure 2-1 1 mW Power Reference Level Test Setup Connection Diagram
Ch A
POWER METER
P-SERIES
POWER METER
NOTE
For rear panel options, the connections will differ from the illustration shown here. Refer to
the connector identification markings on the rear panel for further details