Technical data

Performance Tests 2
N1911A/1912A P-Series Power Meters Service Guide 21
Complete Equipment List
Instrument Critical Specifications Recommended Keysight
Model Number
Alternative Keysight
Model Number
Analyzers
Network analyzer N3383A N3381A
N3382A
8753ES/ET
Counters
Universal counter Frequency: 10 MHz
Gate Time: 10 seconds
53132A 53131A
Meters
Power meter Dual channel
Absolute Accuracy: +
0.5 %
Power Reference Accuracy:
+
0.9 % - (a best capability mea-
surement is required for the
Power Reference Output - the
power level must be accurately
measured, and the uncertainty
of this measurement must also
be known)
E4419B E4419A
Power sensor 2 required Frequency: 50 MHz
Amplitude Range: –70 dBm to
–20 dBm
SWR:
1.15 at 50 MHz
8481D
Power sensor Frequency: 50 MHz
Amplitude Range: –30 dBm to
+20 dBm
SWR:
1.1 at 50 MHz
8482A
Attenuators
20 dB
Fixed attenuator
Type-N(m,f) 8491A
(Option 020)
30 dB
Fixed attenuator
Type-N(m,f) 11708A
Miscellaneous Devices
10 MHz
Frequency standard