Technical data
Specifications 1
N1911A/1912A P-Series Power Meters Service Guide 11
Sensor Calibration Uncertainty
9
Definition: Uncertainty resulting from non- linearity in the sensor
detection and correction process. This can be considered as a combination
of traditional linearity, cal factor and temperature specifications and the
uncertainty associated with the internal calibration process.
9
Beyond 70 % Humidity, and additional 0.6 % should be added to these values.
Physical characteristics
Frequency band N1921A N1922A
50 MHz to 500 MHz 4.5 % 4.3 %
500 MHz to 1 GHz 4.0 % 4.2 %
1 GHz to 10 GHz 4.0 % 4.4 %
10 GHz to 18 GHz 5.0 % 4.7 %
18 GHz to 26.5 GHz 5.9 %
26.5 GHz to 40 GHz 6.0 %
Dimensions
N1921A
N1922A
135 mm x 40 mm x 27mm
127 mm x 40 mm x 27 mm
Weights with cable
Option 105
Option 106
Option 107
0.4 kg
0.6 kg
1.4 kg
Fixed sensor cable lengths
Standard
Option 106
Option 107
1.5 m (5-feet)
3.0 m (10-feet)
10 m (31-feet)