Technical data

Specifications 1
N1911A/1912A P-Series Power Meters Service Guide 9
6
Internal trigger latency is defined as the delay between the applied RF crossing the trigger level
and the meter switching into the triggered state.
7
External trigger latency is defined as the delay between the applied trigger crossing the trigger
level and the meter switching into the triggered state.
8
External trigger output latency is defined as the delay between the meter entering the triggered
state and the output signal switching.
Trigger delay
Delay range
± 1.0 s, maximum
Delay resolution
1% of delay setting, 10 ns maximum
Trigger hold-off
Range
1
μ
s to 400 ns
Resolution
1% of selected value (to a minimum of 10 ns)
Trigger level threshold hysteresis
Range
± 3 dB
Resolution
0.05 dB