1 Specifications
8 N1911A/1912A P-Series Power Meters Service Guide
P-Series Power Meter Specifications
Meter uncertainty
Instrumentation linearity ± 0.8 %
Timbebase
Timebase range 2 ns to 100 msec/div
Accuracy ± 10 ppm
Jitter
≤
1
ns
Trigger
Internal trigger
Range
–20 to +20 dBm
Resolution
0.1 dB
Level accuracy
± 0.5 dB
Latency
6
160 ns ± 10 ns
Jitter
≤
5 ns rms
External TTL trigger input
High
> 2.4 V
Low
< 0.7 V
Latency
7
90 ns ± 10 ns
Minimum trigger pulse width
15 ns
Minimum trigger repetition period
50 ns
Impedance
50
Ω
Jitter
≤
5 ns rms
External TTL trigger output
Low to high transition on trigger event.
High
> 2.4 V
Low
< 0.7 V
Latency
8
30 ns ± 10 ns
Impedance
50
Ω
Jitter
≤
5 ns rms