Technical data

6Repair Guide
108 N1911A/1912A P-Series Power Meters Service Guide
Additional Repair Notes
Replacing A Sensor Flex Assembly:
• The Sensor flex assembly is supplied straight
• Create A sharp bend (Figure 6- 20): The flex circuit must be bent at a
right- angle where it meets the printed circuit board. It can only be bent
after heat has been applied to it (i.e. using a hot- air gun, or a similar
device)
Figure 6-20 Creating a sharp bend
NOTE
Once this sharp bend has been created, the flex should not be bent at this point again; to
do so may break the tracking within the flex.
Route and connect the sensor flex assembly: once the sensor flex assembly has been
attached to the power meter; it should be folded to match the route taken by the
assembly being replaced. Heat may be used to assist the folding of the flex.