Technical data

Repair Guide 6
N1911A/1912A P-Series Power Meters Service Guide 101
Step 7:
Remove the 3 screws that attach the
calibrator assembly to the display support
molding, and separate them from one
another
Note:
Take care not to damage the EMI spring
fingers on the calibrator assembly
Step 8:
Disconnect the backlight cable assembly
from the display interface board and inverter
board
Step 9:
Remove the 2 screws that attach the display
inter face board to the inverter board, and
separate them from one another
Instructions Visual
EMI Spring
Finger