Technical data

Repair Guide 6
N1911A/1912A P-Series Power Meters Service Guide 99
Step 2:
Use ODU socket to remove N1912-61806
sensor flex assembly
Release the tab holding the flex cable to the
display interface board, and then disconnect
it
Step 3:
Disconnect the white plastic plug from the
display interface board
Step 4:
Unlock the main plastic clip that holds the
front panel sub-frame and display support
molding together (situated beside the key
flex circuit), and carefully pull them apart to
separate them
Instructions Visual
Flex Cable
White Plastic Plug