Technical data

6Repair Guide
96 N1911A/1912A P-Series Power Meters Service Guide
DAP assembly, Channel A (Rear
connectors option)
Remove handle, bumpers, and top clamshell
(including PSU)
Disconnect the sensor RF connections from the
main board
Remove the 4 screws securing the DAP assembly to
the main board
DAP assembly can now be removed
DAP assembly, Channel B Remove handle, bumpers, and top clamshell
(including PSU)
Remove the 4 screws securing the DAP assembly to
the main board
DAP assembly can now be removed
Front panel assembly (front connectors
option)
Remove handle, bumpers, and top clamshell
(including PSU)
Disconnect the sensor RF connections from the
main board
Disconnect the sensor flex connection(s) from the
main board
Disconnect the calibrator assembly cable
connection from the main board
Disconnect the EMI earth wires from the calibrator
assembly
Disconnect main board ribbon cable from the front
panel
Front panel assembly can now be removed
Front panel assembly (rear
connectors option)
Remove handle, bumpers, and top clamshell
(including PSU)
Disconnect calibrator semi-rigid from the rear panel
assembly
Disconnect the calibrator assembly cable
connection from the main board
Disconnect the EMI earth wires from the calibrator
assembly
Disconnect main board ribbon cable from the front
panel
Front panel assembly can now be removed
Sensor flex assembly (front connectors
option)
[Remove front panel assembly as previously
described]
Use the N1912-61807 special tooling kit to remove
the sensor flex assembly
Disassembly of Replacement Part Instructions