Technical data

6Repair Guide
92 N1911A/1912A P-Series Power Meters Service Guide
Reassembly Instructions
Instructions Visual
The reassembly process is simply the
reverse of the disassembly process.
However, there are various points to be
aware of:
USB/LAN connectors must rest on top of
the rear panels’ EMC spring fingers.
The position of the cable clamp depends
on whether option 101 or 003 is fitted.
The main board connector from the PSU
must be pushed firmly to fully engage it.
Take care not to trap any cables when
fitting the top clamshell.
Analog recorder output connections
(Figure 6-15): Ensure recorder 1 is plugged
into the rear connector. Where applicable,
recorder 2 is plugged into the connector
nearer the front.
Figure 6-15
Sensor flex connections (Figure 6-16):
A – Front, Channel A (for Option 101)
B – Front, Channel B (for Option 101)
C – Rear, Channel A (for Option 003)
D – Rear, Channel B (for Option 003)
Figure 6-16
A
B
C
D