Technical data

6Repair Guide
88 N1911A/1912A P-Series Power Meters Service Guide
Disconnect the front panel cable
(Figure 6-9): Depress both sides of
theconnector holding the ribbon cable to
eject it.
Figure 6-9
Disconnect the semi-rigid cable
(Figure 6-10):
Note:
This only applies to Option 003 units.
Use the 5/16” spanner to disconnect the
semi-rigid cable from the N-Type connector on
the rear panel.
Figure 6-10