Technical data

Repair Guide 6
N1911A/1912A P-Series Power Meters Service Guide 87
With reference to Figure 6-6 and Figure 6-7:
Lift and remove the cable clamp.
Disconnect the cable attaching the fan
assembly to the main board.
Lift and remove the fan assembly
Disconnect the sensor RF connections from
the main board.
Disconnect the sensor flex connection(s)
from the main board.
Disconnect the calibrator cable connection
from the main board.
Disconnect the analog recorder output
connection(s) from the main board.
Figure 6-7
Remove the EMI earth wires (Figure 6-8:
Use the 9/32” Socket to remove the hex nut
attaching the EMI earth wires to the
calibrator assembly. Remove the earth
wires and washers, taking note of the
assembly order.
Figure 6-8
Instructions Visual
1
2
3
6
4
5
7
8
9
11
10
12
1314