Technical data

6Repair Guide
84 N1911A/1912A P-Series Power Meters Service Guide
Disassembly Instructions
The guidelines in this section describe the disassembly of the major
assembling in the Keysight N1911A and N1912A power meters.
Instructions Visual
This procedure focuses primarily on model
N1912A, Option 101 (i. e. dual channel, with
front panel sensor and power reference
connectors)
Additional information is provided to assist
in the disassembly of Option 003 units (i. e.
with rear panel sensor and power reference
connectors)
Remove the handle: Rotate it to the vertical
position. Pull both sides outwards from the
body of the unit.
Remove the front/rear bumpers: Pull one
side of the bumper outwards to disengage
it. Pull it away from the unit.
Separate the clamshells (Figure 6-2): Use
the T20 Torx screwdriver bit to loosen the 4
captive screws.
Figure 6-2
N1912A,
Option 101
N1911A,
Option 003
LXI