Technical data

SERVice Subsystem 16
N1911A/1912A P-Series Power Meters Programming Guide 707
SERVice:BIST:TBASe:STATe <boolean>
This command sends a 10 MHz time base signal to the rear panel trig out
for testing purposes.
If the command is set to:
ON, the 10 MHz time base signal is sent to the rear panel trigger out
connector.
OFF, the 10 MHz time base signal is disabled.
Syntax
Example
NOTE
This command overrides the OUTPut:TRIGger[:STATe] command.
For example, if OUTPut:TRIGger[:STATe] is ON and the command
SERV:BIST:TBAS ON is sent, this command overrides the Trigger state and sets it to
OFF. However, the 10 MHz remains out the Trig out port.
If the SERV:BIST:TBAS ON has been sent, the 10 MHz is on and the
OUTPut:TRIGger[:STATe] is then toggled to ON, the channel trigger is now routed to
the Trig out overriding the service command turning the 10 MHz to off.
SERV:BIST:TBAS:STAT OFF This command disables the signal.
:BIST
SERV
0|OFF
1|ON
Space
?
:TBAS