Technical data

TRACe Subsystem 13
N1911A/1912A P-Series Power Meters Programming Guide 621
present, otherwise Error –241, "Hardware Missing" is generated.
If free run acquisition is selected or sensor average mode selected,
Error –221 "Settings Conflict" occurs.
NOTE
If you attempt to measure a pulse out of the range of the capture, for example, measure the
5th pulse and there are only 4 pulses displayed, the power meter returns #0##9.91E37 as
the result.