Technical data

11 STATus Subsystem
474 N1911A/1912A P-Series Power Meters Programming Guide
set when a power sensor is disconnected.
STATus:DEVice:PTRansition is 1, then STATus:DEVice:EVENt? is
set when a power sensor is connected.
The Channel A and B sensor error bits (3 and 4) are set to:
1, if the P- Series, N8480 Series or E- Series power sensor EEPROM has
failed or if there are power sensors connected to both the rear and
front panel connectors.
0, for every other condition.
The Front Panel key press bit (bit 14), when queried with the
STATus:DEVice:EVENt? query indicates whether any front panel keys
have been pressed since power up or since you last queried the device
status register. This bit ignores the :NTRansition, and :PTRansition
registers and a :CONDition? query always returns a 0.
NOTE