Technical data

SENSe Subsystem 10
N1911A/1912A P-Series Power Meters Programming Guide 437
Example
Query
[SENSe[1]]|SENSe2:SWEep[1]|2|3|4:AUTO:REF1|REF2?
The query returns the current setting of Auto Gating Marker Reference 1
or 2 for the selected gate in numerical value.
Query Example
Error Messages
If no sensor or wrong sensor is connected to the channel, error –241
"Hardware missing" occurs.
If channel is not in trigger acquisition mode, error –221 "Settings
conflict" occurs.
If limits of the values keyed in are exceeded, error –222 "Data out of
range; upper (or lower) limit exceeded; no change" occurs.
SENS1:SWE1:AUTO:REF1 10.0 This command sets the Channel A Auto
Gating Marker Reference 1 to 10 % for
Gate 1.
SENS2:SWE2:AUTO:REF2 40.0 This command sets the Channel B Auto
Gating Marker Reference 2 to 40% for
Gate 2.
SENS1:SWE2:AUTO:REF1? The query returns the current Gate 2
Reference 1 value of Auto Gating Marker.