Technical data
SENSe Subsystem 10
N1911A/1912A P-Series Power Meters Programming Guide 435
Error Messages
• If no sensor or wrong sensor is connected to the channel, error –241
"Hardware missing" occurs.
• If channel is not in trigger acquisition mode, error –221 "Settings
conflict" occurs.
• If Auto Gate fails, error –221 "Settings conflict; Auto Once failed"
occurs.