Technical data

10 SENSe Subsystem
420 N1911A/1912A P-Series Power Meters Programming Guide
Example
Query
[SENSe[1]]|SENSe2:FREQuency[:CW|:FIXed]:STEP?
This query is used to retrieve the number of steps in average or peak
frequency sweep.
Query Example
On Reset
On *RST, the value is set to 0.
Error Messages
If no sensor or wrong sensor is connected to the channel, error –241,
“Hardware missing” occurs.
If E4410, N8480 (excluding Option CFT), E9300, E9320 or N1920 sensor
is connected but acquisition mode is in free run, error –221, “Setting
conflict. Invalid acquisition mode” occurs.
If parameter set is lower than 0, error –222, “Data out of range; value
clipped to lower limit” occurs.
FREQ:STEP 10 This command sets frequency sweep with
10 steps for Channel A.
FREQ:STEP? This query returns the number of steps in
frequency sweep for Channel A.