Technical data

SENSe Subsystem 10
N1911A/1912A P-Series Power Meters Programming Guide 375
Reset Condition
On reset, the filter length is set to 4.
Query
AVERage2:COUNt?
The query returns the current setting of the video filter length. The format
of the response is <NR1>.
Query Example
Error Messages
If the command is used when a sensor other than a P- Series or E9320
power sensor is connected, error –241, “Hardware missing” occurs
If the command is used when an E9320 sensor is connected and set to
AVERage mode rather than NORMal mode, the error –221, “Settings
Conflict” occurs.
AVER2:COUN? This command queries the video filter
length for Channel A.