Technical data

9 PSTatistic Subsystem
348 N1911A/1912A P-Series Power Meters Programming Guide
Query
PSTatistic[1]|2:CCDF:TRACe:[:STATe]?
The query enters a 1 or 0 into the output buffer indicating the status of
the displayed CCDF trace.
1 is returned when there is a trace displayed on the CCDF screen
window
0 is returned when there is no trace displayed on the CCDF screen
window
Query Example
Error Messages
If command is executed in other window besides expanded CCDF
window, error –221 "Settings conflict: Requires CCDF expanded
window" occurs.
If no power sensor is connected, error –241 "Hardware missing" occurs.
If sensor/sensors connected are not P- Series sensors, error –241
"Hardware missing" occurs.
PST:CCDF:TRAC? This command queries whether there is a
trace displayed or not on the CCDF screen
window.