Technical data

PSTatistic Subsystem 9
N1911A/1912A P-Series Power Meters Programming Guide 343
Reset Condition
On reset, the previously saved reference trace will be cleared.
Query
PSTatistic[1]|2:CCDF:STORe:REFerence?
The query enters a 1 or 0 into the output buffer indicating the status of
the CCDF reference.
1 is returned when there is a saved reference trace
0 is returned when there is no saved reference trace
Query Eample
Error Messages
If no power sensor is connected, error –241 "Hardware missing" occurs.
If sensor/sensors connected are not P- Series sensors, error –241
"Hardware missing" occurs.
If measurement speed setting is FAST, error –221 "Settings conflict"
occurs.
If command is executed in other window besides expanded CCDF
window, error –221 "Settings conflict: Requires CCDF expanded
window" occurs.
If the acquisition mode is in continuous triggering or triggering source
is set to either INT1, INT2, or EXT, error –221 "Settings conflict"
occurs.
PST:CCDF:STOR:REF? This command queries whether there is
saved reference trace or not.