Technical data

7MEMory Subsystem
286 N1911A/1912A P-Series Power Meters Programming Guide
Query Example
Error Messages
If more than 81 calibration factors for sensor calibration tables, or 80
offsets for frequency dependent offset tables are in the list, error –108,
“Parameter not allowed” occurs.
If a table is not specified using the MEMory:TABLe:SELect command,
the data cannot be entered and error –221, “Settings conflict” occurs.
If any of the calibration/offset factors are outside of the allowed range,
error –222, “Data out of range” occurs.
MEM:TABL:GAIN? This command queries the calibration
factor/offset in the current table.