Technical data

MEMory Subsystem 7
N1911A/1912A P-Series Power Meters Programming Guide 259
MEMory:CATalog[:ALL]?
This command lists stored sensor calibration tables (8480 Series sensors
and N8480 Series Sensors with Option CFT only), frequency dependent
offset tables and save/recall registers.
The power meter returns the data in the form of two numeric parameters
and as many strings as there are stored tables and save/recall registers:
<numeric_value>,<numeric_value>{,<string>}
The first numeric parameter indicates the amount of memory, in bytes,
used for the storage of tables and registers.
The second numeric parameter indicates the memory, in bytes,
available for the storage of tables and registers.
Each string parameter returned indicates the name, type and size of a
stored table or save/recall register:
<string>, <type>, <size>
<string> indicates the name of the table or save/recall register.
<type> indicates TABL for sensor calibration and frequency
dependent offset tables, or STAT for a save/recall register.
<size> indicates the size of the table or save/recall register in bytes.
A sample of a response may look like the following:
1178,26230,"DEFAULT,TABL,14","8481A,TABL,116",
"8482A,TABL,74",..........."State0,STAT,1619",
"State1,STAT,1619","State2,STAT,1619" ...........
The power meter is shipped with a set of predefined sensor calibration
tables. The data in these sensor calibration tables is based on statistical
averages for a range of Agilent Technologies power sensors. These tables
can be edited. The predefined data is listed in your user’s guide. These
power sensors and table numbers are listed in Table 7- 19.
NOTE
Predefined sensor calibration table is not applicable for N8480 Series power sensors with
Option CFT. Therefore you are required to create a new sensor calibration table for the
sensors when a sensor calibration table is needed.