Technical data

4 16800 Series Portable Logic Analyzers Installation/Quick Start Guide
Features, Logic Acquisition
1 M to 32 M memory depth per channel (depending on
memory option), software upgradeable.
250 MHz or 500 Mb/s maximum state data rate (depending
on state speed option), software upgradeable. The 500 Mb/s
maximum state data rate option is available on the
68-channel and above logic analyzer models.
1 GHz, 64 M deep timing analysis on half channels.
Eye finder (automatic threshold and sample position setup)
feature.
4 GHz timing zoom with 64 K memory depth.
Features, Mainframe
Built-in 15 inch TFT color LCD display, 1,024 x 768 (XGA)
resolution. Touch screen with 16800A Option 103.
80 GB hard disk drive (or external hard drive 16800A Option
109).
10/100 Base-T LAN port.
USB 2.0 ports (six total, two on front, four on back).
One PCI expansion slot.
One PCI Express x1 expansion slot.
Windows® XP Professional operating system.
Agilent Logic Analyzer application which takes the
complexity out of making logic analyzer measurements. You
can perform all operations directly from one window.
Features, Pattern Generator
24 channels at 300 MHz clock; 48 channels at 180 MHz clock.
Memory Depth: 16,777,216 vectors in half-channel mode.
Logic Level (data pods): 5 V TTL, 3-state TTL, 3-state
TTL/CMOS, 3-state 1.8 V, 3-state 2.5 V, 3-state 3.3 V, ECL, 5 V
PECL, 3.3 V LVPECL, and LVDS.
Data Inputs: 3-bit pattern level sensing (clock pod).
Clock Output: Synchronized to output data, delay of 7 ns in
14 steps (clock pod).