Specifications

16
Pattern Generator Specifications and Characteristics
(16821A, 16822A, and 16823A)
Agilent 10476A 3-state 1.8-volt data pod
Output type 74AVC16244
3-state enable Negative true, 38 K to GND, enabled
on no connect
Maximum clock 300 MHz
Skew [1] Typical < 1.5 ns; worst case = 2 ns
Recommended lead set Agilent 10498A
Agilent 10483A 3-state 3.3-volt data pod
Output type 74AVC16244
3-state enable Negative true, 38 K to GND, enabled
on no connect
Maximum clock 300 MHz
Skew [1] Typical < 1.5 ns; worst case = 2 ns
Recommended lead set Agilent 10498A
Agilent E8141A LVDS data pod
Output type 65LVDS389 (LVDS data lines)
10H125 (TTL non-3-state channel 7)
3-state enable Positive true TTL; no connect=enabled
Maximum clock 300 MHz
Skew Typical < 1 ns; worst case = 2 ns
Recommended lead set E8142A
Recommended lead set Agilent 10498A
[1] Typical skew measurements made at pod connector with approximately
10 pF/50 Kload to GND; worst case skew numbers are a calculation of worst
case conditions through circuits. Both numbers apply to any channel within the
pattern generator.
10 K
65LVDS389
ENABLE
3.3 V
LVDS DATA OUT
3-STATE IN TTL
74AVC16244
74AVC16244