Specifications

9
Timing (asynchronous)
analysis mode
Conventional timing Transitional timing
2
Sample rate on all channels 500 MHz 500 MHz
Sample rate in half channel mode 1 GHz N/A
Number of independent analyzers
1
2 (1 for 16801A or 16821A) 2 (1 for 16801A or 16821A)
Sample period (half channel) 1.0 ns N/A
Minimum sample period (full channel) 2.0 ns 2.0 ns
Minimum data pulse width 1 sample period + 1.0 ns 1 sample period + 1.0 ns
Time interval accuracy ± (1 sample period + 1.25 ns + 0.01% of time
interval reading)
± (1 sample period + 1.25 ns + 0.01% of
time interval reading)
Memory depth in full channel mode Option 001: 1 M samples Option 001: 1 M samples
Option 004: 4 M samples Option 004: 4 M samples
Option 016: 16 M samples Option 016: 16 M samples
Option 032: 32 M samples Option 032: 32 M samples
Memory depth in half channel mode Option 001: 2 M samples
N/A
Option 004: 8 M samples
Option 016: 32 M samples
Option 032: 64 M samples
Maximum trigger sequence speed 250 MHz 250 MHz
Maximum trigger sequence levels 16 16
Trigger sequence level branching Arbitrary 4-way if/then/else Arbitrary 4-way if/then/else
Trigger position Start, center, end, or user-defined Start, center, end, or user-defined
1. Independent analyzers may be either state or timing. When the 500 Mb/s state mode is selected, only one analyzer may be used.
2. Transitional timing speed and memory depth are halved unless a spare pod pair (34 channels) is unassigned.
Agilent 16800 Series Logic Analyzer Specifications and Characteristics
(continued)