Specifications

8
State (synchronous) analysis mode Option 250 Option 500
(Available on 16802A, 16803A, 16804A,
16806A, 16822A and 16823A)
Clock qualifier setup time 500 ps N/A
Clock qualifier hold time 0 N/A
Time tag resolution 2 ns 1.5 ns
Maximum time count between stored
states
32 days 32 days
Maximum trigger sequence speed 250 MHz 500 MHz
Maximum trigger sequence levels 16 16
Trigger sequence level branching Arbitrary 4-way if/then/else 2-way if/then/else
Trigger position Start, center, end, or user-defined Start, center, end, or user-defined
Trigger resources 16 patterns evaluated as =, =/, >, ≥, <, ≤
14 double-bounded ranges evaluated as
in range, not in range
1 timer for every 34 channels
2 global counters
1 occurrence counter per sequence level
4 flags
14 patterns evaluated as =, =/, >, ≥, <, ≤
7 double-bounded ranges evaluated as in
range, not in range
1 occurrence counter per sequence level
4 flags
Trigger resource conditions Arbitrary Boolean combinations Arbitrary Boolean combinations
Trigger actions Go To
Trigger, send e-mail, and fill memory
Trigger and Go To
Store/don’t store sample
Turn on/off default storing
Timer start/stop/pause/resume
Global counter increment/decrement/
reset
Occurrence counter reset
Flag set/clear
Go To
Trigger and fill memory
Store qualification Default (global) and per sequence level Default (global)
Maximum global counter 2E+24 N/A
Maximum occurrence counter 2E+24 2E+24
Maximum pattern width Smaller of 128 bits or maximum number of
channels
Smaller of 128 bits or maximum number of
channels
Maximum range width Smaller of 64 bits or maximum number of
channels
Smaller of 64 bits or maximum number of
channels
Timers range 60 ns to 2199 seconds N/A
Timer resolution 2 ns N/A
Timer accuracy ± (5 ns +0.01%) N/A
Timer reset latency 60 ns N/A
Agilent 16800 Series Logic Analyzer Specifications and Characteristics
(continued)