Specifications

18
Clock cable characteristics without a clock pod
The pattern generator clock cables without a clock pod provide an ECL termi-
nated (1 KΩ to –5.2 V) differential signal (from a type 10E164 driver). These are
usable when received by a differential receiver, preferably with a 100 Ω termina-
tion across the lines. These signals should not be used single ended due to the
slow fall time and shifted voltage threshold (they are not ECL compatible).
100 Clock in
10E116
7
8
100 Wait 1, 2, 3 IN
10H125
11, 13, 15
12, 14, 16
–3.25 V
215
215
Clock out
10E164
–3.25 V
Pattern Generator Specifications and Characteristics