Specifications

91
Chapter 1: Agilent Technologies 16700A/B-Series Logic Analysis System
Using the Correlation Dialog
Using the Correlation Dialog
Two required elements of a multiple instrument configuration are the
Arming Signal and the Time Correlation of Data Sets. The
Correlation dialog is used to help you correct any configuration
problems with these two elements. If you run a measurement and the
Correlation dialog appears, reconfigure any inputs that show errors.
Armed From – For each Input Name, set the Armed From field to either
the Group Run or another instrument. This allows the arming signal to
create an ordered triggering sequence between instruments. For more
information, refer to "The Arming Signal", in the Overview - Multiple
Instrument Configuration (see page 67).
Time Tags – Each data set must be acquired in reference to a common
time base or time reference. Timing analyzers are always referenced to the
timebase of the measurement. However, if you are using a State analyzer,
you must set its Count field to Time (time tags). For more information,
refer to "Time Correlation of Data Sets", in the Overview - Multiple
Instrument Configuration (see page 67).