Specifications
90
Chapter 1: Agilent Technologies 16700A/B-Series Logic Analysis System
Using the Mixed Signal Tab
Using the Mixed Signal Tab
A mixed signal measurement captures data and displays results from
more than one instrument tool. All displayed signals are time-
correlated and typically displayed in the same display tool. An example
of a mixed signal application would be when Using a Timing Analyzer
and an Oscilloscope (see page 83).
The Mixed Signal tab is used to import signals from other instruments
into the display tool.
NOTE: Ideally, before you use the Mixed Signal tab to import a signal, the instrument
tool from which the signal will come should be configured and able to
successfully capture its own data set. If any required part of the configuration
is not correct, such as Arming or Time Tags, a Correlation Dialog (see
page 91) will appear to help correct the configuration.
1. Select the Mixed Signal tab.
2. Select the desired signal.
3. Select Connect.
4. Select Group Run. Running the measurement at this point makes the new
imported signals available for insertion.
5. From the menu bar in the display window, select Edit, then select Insert.
6. From the Label Selection Dialog that appears, select the desired signal to
import, then select Apply.
7. Select OK.