Specifications

72
Chapter 1: Agilent Technologies 16700A/B-Series Logic Analysis System
Overview - Multiple Frames Configuration
The Multi-frame configuration dialog is used to identify all frames
included in the Multi-frame group. A Clear Multi-frame Run (see
page 82) field is available if you need to stop and resynchronize all
frames. A Telnet (see page 142) field is also available to quickly
configure a Telnet connection to other instruments.
Enter the host names of all frames used in the Multi-frame
configuration. If the frame has already been assigned a host name, it
will appear in the main System window title bar. If a host name has not
been assigned, ask your System Administrator to assign one.
NOTE: All frames that are cabled together by a Multi-frame cable must be entered
into this table in order to guarantee proper operation of any of the frames.
Disconnect the Multi-frame cable from any frames not listed in the host table.
2. Intermodule
Configuration
The Intermodule window is accessed by selecting the Intermodule
Icon from the Icon Bar at the top of the main System window. Use the
Intermodule window to combine multiple measurement modules
together to run as a group. They are configured in a Group Run
Arming Tree to coordinate the order in which each module is armed to
perform its part of the measurement. The benefit of a Multi-frame
intermodule measurement is that you can capture multiple data sets
from any frame, then view the data sets with time correlated global
markers.
In all Intermodule measurements, as each tool is selected from an
Independent status, you have the choice to arm this tool by other