Specifications

30
Chapter 1: Agilent Technologies 16700A/B-Series Logic Analysis System
System Overview
procedures are traceable and must specify adequate calibration
standards.
Calibration procedures verify products meet the specifications by
comparing measured parameters against a pass-fail limit. The pass-fail
limit is the specification less any required guardband.
The term "calibration" refers to the process of measuring parameters
and referencing the measurement to a calibration standard rather than
the process of adjusting products for optimal performance.
What is a Function Test
Function tests are quick tests designed to verify basic operation of a
product. Function tests include operator's checks and operation
verification procedures. An operator's check is normally a fast test
used to verify basic operation of a product. An operation verification
procedure verifies some, but not all, specifications, and often at a lower
confidence level than a calibration procedure.
How the Help System is Accessed
The Agilent 16700A/B-series logic analysis systems are modular.
Agilent offers a choice of hardware and software tools you may install.
As you add new tools, you also add the tool's help volume into the
system.
When Help is accessed from the menu bar in the main System window,
you get the system help volume. If you access Help from any analyzer,
oscilloscope, pattern generator, or display/analysis tool window, you
get help for that specific tool.