Specifications
16
Chapter 1: Agilent Technologies 16700A/B-Series Logic Analysis System
Using the Target Control Port
4. Set Output Drive to Open Collector.
5. When the target needs to be reset manually, select Pulse.
About Open Collector and Active Drive
Open Collector functions as a tri-state, with a logic "1" being high-
impedance, and a logic "0" sinking a current of up to 12 milliamps.
Active Drive puts out a standard TTL signal with 1 high and 0 low.
Both of these signal types must be properly terminated to function
properly. This is especially important for edge-sensitive circuitry.
About Toggle and Pulse
Toggle will flip the settings of the signal and leave them that way. For
example, if your signal is set to 1 and you select Toggle, the setting
changes to 0.
Pulse flips the settings for one clock cycle, at least 16 ms. The pulse
may last longer. You cannot specify the duration of the pulse.