Specifications
113
Chapter 1: Agilent Technologies 16700A/B-Series Logic Analysis System
Run/Group Run Function
Run/Group Run Function
Using Run - Run All - Group Run
The Run/Stop functions are initiated by selecting icons in the icon bar
at the top of the tool windows. All instrument, display, and analysis tool
windows will have one of the Run icons shown below to initiate the run
function.
When two or more instrument tools are configured, they can be run
either independently or as a group. If run in a group, it is called an
Intermodule measurement. Use the Intermodule Window (see the
Agilent Technologies 16700A/B-Series Logic Analysis System help
volume) to coordinate the arming in a "Group Run". A common "Group
Run" configuration is to configure one instrument to trigger and then
arm another instrument to start evaluation of its own trigger condition.
The Run Single icon appears if you have a single instrument
configured in your measurement and you want to run a single
acquisition.
The Run All icon always appears in the System, Workspace and
Run Status windows. Also appears in instrument and display windows
when you are using multiple instruments in your measurement and
these instruments ARE NOT configured in an intermodule
measurement (Group Run). This choice runs a single acquisition on all
instruments in the configuration.
The Group Run icon appears in all windows when you are using
multiple instruments, and these instruments are configured into a
Group Run. This choice runs a single acquisition on all instruments in
the Group Run configuration.
The Run Repetitive icon appears in all windows. It is used to run a
Run Single, Run All, and a Group Run acquisition repetitively. The
current run mode will continue to run until Cancel is selected.