Specifications
104
Chapter 1: Agilent Technologies 16700A/B-Series Logic Analysis System
Adjusting Intermodule Skew
Trigger Setup
1. From the bottom of the oscilloscope window, select the Trigger field.
2. Set the Trigger Mode field to trigger Immediate.
This will cause the oscilloscope to trigger and begin running as soon as the
arming signal is received from the Timing analyzer.
Configure the Group Run Arming Tree
1. From the Intermodule window, select the analyzer icon, then select Group
Run as the arming device.
2. Select the oscilloscope icon, then select the analyzer as the arming device.
The result is an arming tree that has the analyzer armed when the Group
Run field is selected. Then the scope is armed, when the analyzer triggers
on the glitch, and the scope is configured to trigger immediately.