User`s guide

3-18 Agilent 4155C/4156C User’s Guide Vol.2, Edition 6
Sampling Measurements
Measurement Functions
Table 3-2 Effective Parameter Values
a
Number of units (SMUs or VMUs) used for measurements.
b
If voltage source SMU is connected to R-BOX, only this SMU can be used for the
measurement.
c
Maximum 8 units if the 41501 is used (6 SMUs and 2 VMUs). If multiple
measurement units are used, total measurement time will be more than the sum of
the measurement time by all units.
d
If compliance value is more than the measurement range, range change when
starting measurement makes long measurement time. Also if ranging mode is set
to AUTO or LIMITED, range changing makes long measurement time.
e
Integration time is an element of the measurement time. If automatic measurement
data compensation is executed, the measurement time will be more than two times
the integration time.
INITIAL INTERVAL
720 μs to 1.92 ms 2 ms to 65.535 s
NO.OF SAMPLES
Max. 10001/(number of measurement units)
TOTAL SAMP.TIME
NO LIMIT or
INITIAL INTERVAL × (NO.OF SAMPLES 1) s
to 1 × 10
11
s
HOLD TIME
30 ms to 655.35 s,
100 μs resolution
0to655.35 s,
100 μsresolution
Stop Condition DISABLE DISABLE/ENABLE
Measurement Units
a
1
b
Max. 8
c
Measurement Range
d
FIX FIX/AUTO/LIMITED
Integration Time
e
Short Short/Medium/Long