Sampling Measurements
Measurement Functions
Source Output Sequence and Time Origin
Source unit output sequence and the time origin depends on the setup value of the
OUTPUT SEQUENCE MODE OF SAMPLING field in the MEASURE: OUTPUT
SEQUENCE screen. The following two modes are available for the field.
SIMULTANEOUS mode
All source unit starts output at same timing. This timing is defined as the Time
Origin. See figure below.
CONST
CONST
Pulse Bias
CONST
Hold time
Time origin
Sampling start Sampling completion
SMU2
SMU3
PGU1
SMU1
0 V
0 V
0 V
0 V