Agilent 4155C/4156C Users Guide Vol.2, Edition 6 2-15
Knob Sweep Measurements
Executing Measurements
To copy knob sweep setups
Select SETUP COPY primary softkey. This copies knob sweep setups to the
MEASURE: SWEEP SETUP and DISPLAY: DISPLAY SETUP screens. Then you
can see the setups on the screens, and use the setups for the normal sweep
measurement.
Note that the power compliance function is not available for the knob sweep
measurement. So the POWER COMP column is set to OFF.
Softkey Description
VAR2 SETUP Used to change VAR2 sweep source setup. The
following secondary softkeys are available:
VAR2 START Sets VAR2 sweep start value.
VAR2 STEP Sets VAR2 sweep steps.
VAR2 POINTS Sets number of sweep steps.
COMPLIANCE Sets VAR2 compliance value.
CONST SETUP Used to change CONST source setup. Secondary
softkeys are available for selecting CONST source.
Select a secondary softkey to change the CONST
source output value.