User`s guide
Agilent 4155C/4156C User’s Guide Vol.2, Edition 6 1-41
Sweep Measurements
Making a Measurement
Step 6. Set up the results display
You set the results display information on the DISPLAY: DISPLAY SETUP screen.
1. Press Display front-panel key. The DISPLAY: DISPLAY SETUP screen appears.
2. Make sure GRAPHICS is displayed in the DISPLAY MODE field. If not, select
GRAPHIC secondary softkey in the DISPLAY MODE field.
3. Set the X-, Y1-, and Y2-axes information as shown below. For the actual setup,
see Figure 1-9.
on Front Panel on Keyboard
To enter "VG"
in NAME field,
select VG secondary
softkey.
press
Shift-F3 keys.
To set "LINEAR"
in SCALE field,
select LINEAR
secondary softkey.
press
Shift-F1 keys.
To enter "0.00000 V"
in MIN field,
press
0, then Enter. type 0, then press Enter.
To enter "2.00000 V"
in MAX field,
press
2, then Enter. type 2, then press Enter.
To enter "SQID"
in NAME field,
select MORE 1/2, then
SQID secondary
softkeys.
press
Shift-F7 keys, then
Shift-F3 keys.
To enter "0.00000 A"
in MIN field,
press
0, then Enter. type 0, then press Enter.
To enter "100.000mA"
in MAX field,
press
1, 0, 0, m, then Enter. type 100m, then press
Enter.
To enter "DSQID"
in NAME field,
select MORE 1/2, then
DSQID secondary
softkeys.
press
Shift-F7 keys, then
Shift-F4 keys.